3. Sterilization
For the sterilization method, we recommend the autoclave sterilization method.
Sterilization is required first time you use and after each patient as noted below.
Autoclaving
1
)
Turn off the Power.
2
)
Detach the motor from the Motor Cord.
(
Refer to 2-1 Connecting/disconnecting
the Motor and the Motor Cord
)
3
)
Clean the surface of the Motor with brush.
(
Do not use the metal brush
)
, and wipe
it with the cotton moistened with disinfecting alcohol.
4
)
Screw the Motor Cap to the Motor. Put the Autoclave plug to the motor insert.
5
)
Insert into an autoclave pouch. Seal the pouch.
6
)
Autoclavable up to max. 135°C.
ex.
)
Autoclave for 20 min. at 121°C, or 15 min. at 132°C.
7
)
Keep the Motor in the autoclave pouch to keep it clean until you use it.
Sterilization at 121°C for more than 15 minutes is recommended by EN13060.
Fig. 3
Motor Cap
Autoclave
Plug
O-Ring