OPERATION MANUAL
Please read this Operation Manual carefully
before use, and file for future reference.
OM-E0387E 002
Micromotor for Laboratory Work

Table of Contents

Questions and Answers

  • K
    Kevin FoleySep 23, 2025
    Why does the bur come out of my NSK Laboratory Equipment?
    • J
      James WilliamsSep 23, 2025
      Tighten the chuck securely.
  • T
    Todd SolisAug 23, 2025
    What to do if the motor and handpiece do not run, or Reset Lamp lights on NSK Laboratory Equipment?
    • T
      Tiffany GibsonAug 23, 2025
      If you have multiple units, try replacing the motor cord. If normal operation resumes, the motor or motor cord may be shorted and need repair. If the issue persists, the control unit may require repair.
  • S
    shawncastroAug 25, 2025
    What to do if error code E0 appears on NSK Laboratory Equipment?
    • A
      Amber CallahanAug 25, 2025
      First, try turning the power off and then on again. If the same error code appears again, it indicates that the control unit requires repair.
  • U
    umartinezAug 27, 2025
    What to do if error code E3 appears on NSK Laboratory Equipment?
    • K
      Kathy GrayAug 27, 2025
      First, ensure the motor cord is correctly connected. If the error persists, maintenance mode (3) “HL”: Confirm whether you work by a motor signal check normally. If problems are found during the check, the motor cord might be severed, or the motor sensor could be faulty, requiring repair. Also, ensure the chuck is locked. If the rotation is abnormal, it might require repair of the motor and handpiece.
  • K
    Kelly GrayAug 27, 2025
    What to do if error code E6 appears on NSK Laboratory Equipment?
    • S
      sarahjohnsonAug 27, 2025
      Ensure the chuck is locked. If problems persist during a motor signal check in Maintenance mode (3) “HL”, the motor cord may be severed, or the sensor in the motor may be faulty and require repair. If rotation is abnormal when turning a bur with a finger, the motor and handpiece may need repair. If it work normally, error indication is a temporary thing by the overload.
  • P
    Patrick CantuSep 18, 2025
    What to do if the ball bearings are worn on NSK Laboratory Equipment?
    • D
      Daniel ArroyoSep 18, 2025
      Send the equipment to your dealer for repair.
  • T
    tstewartSep 15, 2025
    How to fix heavy run out of the bur on NSK Laboratory Equipment?
    • S
      Shannon DavidsonSep 15, 2025
      Clean the inside of the chuck and spindle to remove any stuck dust.
  • S
    Shelby LeeSep 16, 2025
    What to do if the chuck is worn on NSK Laboratory Equipment?
    • K
      Krystal WallSep 16, 2025
      Replace the worn chuck.
  • A
    Amanda WalkerSep 6, 2025
    What to do if error code EF appears on NSK Laboratory Equipment?
    • D
      davisrachelSep 6, 2025
      If the Foot Pedal does not operate normally, replace the Foot Pedal or ask for repair.
  • T
    thompsonjulieSep 3, 2025
    How to fix NSK Laboratory Equipment pilot lamp when it does not light?
    • V
      vwrightSep 4, 2025
      Ensure the power plug is correctly inserted. If the pilot lamp still doesn't light, replace the fuse with a specified fuse. If the reason for the blown fuse is unknown, request an inspection. If the power switch is faulty, request a repair.

Summary

GROUNDING INSTRUCTIONS

OTHER WARNING INSTRUCTIONS

Important Instructions and Warning on ULTIMATE XL.

Set up of Control Unit

Mounting of Motor

Connect the motor cord plug locator into the Motor Connector, aligning with the groove, and tighten.

Mounting of Foot Pedal

Insert the Foot Pedal cord plug locator into the Foot Pedal Connector, aligning with the groove.

Connecting of Power Cord

Securely insert the power cord plug into the connector at the back of the unit.

Operation Procedure

HAND Operation

Describes the procedure for operating the micomotor using the hand control.

Speed limit mechanism

Protective Circuit for Motor

Error Code

Replacement of Fuse

Maintenance Mode

Vacuum-coupled Mode

Handling of Motor and Handpiece

Specification

Troubleshooting

NSK Ultimate XL Specifications