What to do if error code EF appears on NSK Laboratory Equipment?
- DdavisrachelSep 6, 2025
If the Foot Pedal does not operate normally, replace the Foot Pedal or ask for repair.
What to do if error code EF appears on NSK Laboratory Equipment?
If the Foot Pedal does not operate normally, replace the Foot Pedal or ask for repair.
What to do if error code E8 appears on NSK Laboratory Equipment?
If you have two or more units, try replacing the motor cord. If the equipment operates normally after replacing the motor cord, the motor or the motor cord may be shorted, requiring repair. If the same error code persists after replacing the motor cord, the Control Unit likely needs repair.
How to fix NSK Laboratory Equipment pilot lamp when it does not light?
Ensure the power plug is correctly inserted. If the pilot lamp still doesn't light, replace the fuse with a specified fuse. If the reason for the blown fuse is unknown, request an inspection. If the power switch is faulty, request a repair.
What to do if error code E5 appears on NSK Laboratory Equipment?
Try turning the power off and on again, repeating the start and stop actions several times. If normal operation resumes, there is no problem. If the same error code appears, the Control Unit may require repair.
What to do if error code E4 appears on NSK Laboratory Equipment?
First, ensure the Vacuum-coupled Mode is off if you're not using a dust collector. If the issue continues, allow the unit to cool down for about 10 minutes, then check operation again. If it operates normally, no problem exists. However, check the operating environment and storage for high temperatures. If the error code appears frequently, the Control Unit may require repair.
What to do if error code E2 appears on NSK Laboratory Equipment?
If you have multiple units, try replacing the motor and motor cord. If normal operation resumes, the motor or motor cord may be severed and require repair. If the error code persists after replacement, the Control Unit may require repair.
What to do if error code E1 appears on NSK Laboratory Equipment?
First, turn the power off and on again. If the equipment operates normally, the error was likely temporary due to an overload, which isn't a problem. If you have multiple units, try replacing the motor and motor cord to check operation. If normal operation resumes, the motor or motor cord may be shorted and need repair. If the same error code appears after replacing, the Control Unit may need repair. Also, check that the motor cord is correctly connected.
What to do if error code E6 appears on NSK Laboratory Equipment?
Ensure the chuck is locked. If problems persist during a motor signal check in Maintenance mode (3) “HL”, the motor cord may be severed, or the sensor in the motor may be faulty and require repair. If rotation is abnormal when turning a bur with a finger, the motor and handpiece may need repair. If it work normally, error indication is a temporary thing by the overload.
What to do if error code E3 appears on NSK Laboratory Equipment?
First, ensure the motor cord is correctly connected. If the error persists, maintenance mode (3) “HL”: Confirm whether you work by a motor signal check normally. If problems are found during the check, the motor cord might be severed, or the motor sensor could be faulty, requiring repair. Also, ensure the chuck is locked. If the rotation is abnormal, it might require repair of the motor and handpiece.
What to do if error code E0 appears on NSK Laboratory Equipment?
First, try turning the power off and then on again. If the same error code appears again, it indicates that the control unit requires repair.
| Brand | NSK |
|---|---|
| Model | Ultimate XL |
| Category | Laboratory Equipment |
| Language | English |
Connect the motor cord plug locator into the Motor Connector, aligning with the groove, and tighten.
Insert the Foot Pedal cord plug locator into the Foot Pedal Connector, aligning with the groove.
Securely insert the power cord plug into the connector at the back of the unit.
Describes the procedure for operating the micomotor using the hand control.