32
English
1. Indications for Use
Osseo 100 is indicated for measuring the stability of
dental implant.
The intended patient population is patients having dental 
implants.
The indication is patients undergoing dental implant 
procedures.
Osseo 100 is contraindicated for implant systems to 
which the MulTipeg could not be attached for mechanical 
incompatibility reasons.
The direct clinical benefit is that the clinician can get an 
objective value (an ISQ-value) of the implant stability. 
2. Intended users
Professional health care users and professional health care 
facility environments only.  
Please read the instruction for use before the first usage.
3. Figures and System components
Fig 1 Osseo 100 Instrument 
Included in package
Fig 2 MulTipeg Driver 
Included in package
Fig 3 Example MulTipeg 
Not included, sold separately
Fig 4 Mains adapter and plugs 
Included in package
Fig 5 Measurement position 
Shows how the instrument tip is held towards the 
MulTipeg during a measurement
Fig 6 ISQ Tester 
Not included, sold separately
Only original parts 
should be used.
Power supply:  
Use only the 
supplied mains 
adapter and plugs.
No user 
modification of 
this equipment is 
allowed.
Batteries should 
be collected 
separately.
4. Specifications
•  Power input: 5VDC, 1 VA
•  Charger input: 100-240 VAC, 5VA
•  Instrument weight: 78g
•  Dimensions instrument: 202mm x 29mm x 25mm
•  Charger safety class: EN 60601-1 Class II
•  Instrument safety class: EN 60601-1 ME Class II
•  EMC: EN 60601-1-2, class B
•  The instrument is intended for continuous use
•  The instrument contains NiMH batteries
•  Battery type: AAA, rechargeable
•  Voltage: 1.2 V
•  Current: 900 mAh
5. Operating environment
Ambient temperature: 16° to 40° C (60°–104° F).  
Relative humidity: 10% to 80% Rh. 
Atmospheric pressure: 500 hPa – 1060 hPa (0.5–1.0 atm).
6. Transport and storage
Ambient temperature: -20° to 40° C (-4°–104° F).  
Relative humidity: 10%–85% Rh.  
Atmospheric pressure: 500 hPa–1060 hPa (0.5–1.0 atm).