-22-
Limitations of warranty
Solely for the benefit of the original buyer/user, Nutricia Medical Devices B.V, warrants all new Flocare
®
Infinity
pumps, of its manufacture to be free from defects in
material and workmanship, excluding normal wear and tear, and will replace or repair, at its service facility or other location designated by Nutricia Medical Devices
B.V, any Flocare
®
Infinity
pump returned to it within thirty-six (36) months of original purchase by the buyer/user. Such repair or replacement shall be free of charge.
Nutricia Medical Devices B.V warrants to the original buyer/user, all repaired or replaced pumps to be free from defects in material and workman-
ship and will replace or repair such products, at its service facility or other location designated by Nutricia Medical Devices B.V. Such repair or replacement shall
carry a warranty of ninety (90) days from the date of repair or replacement or the balance of the new pumps warranty as described above, whichever is greater.
THIS WARRANTY APPLIES ONLY TO FLOCARE
®
INFINITY
PUMPS MANUFACTURED BY NUTRICIA MEDICAL DEVICES B.V AND IS THE ONLY
WARRANTY GIVEN WITH RESPECT TO THE PUMPS. NO WARRANTIES IMPLIED IN LAW, INCLUDING, BUT NOT LIMITED TO THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE, SHALL APPLY. NUTRICIA MEDICAL DEVICES B.V WILL BE
LIABLE, IN ANY EVENT, ONLY FOR THE PURCHASE PRICE OF THE DEFECTIVE PRODUCT, BUT NOT FOR ANY CONSEQUENTIAL DAMAGES.
This Warranty may not be modified, amended or otherwise changed, except by a written document properly executed by a corporate officer of Nutricia Medical Devices B.V
THE WARRANTY IS VOID IF THE FLOCARE
®
INFINITY
PUMP IS SUBJECT TO ABUSE, ACCIDENT, ALTERATION, MODIFICATION, TAMPERING, MISUSE OR THE
UNAUTHORIZED REPAIR OR SERVICE IN ANY WAY WITHOUT PRIOR AUTHORIZATION FROM NUTRICIA MEDICAL DEVICES B.V. IN ANY EVENT, NUTRICIA MEDICAL
DEVICES B.V.’S LIABILITY SHALL NEVER EXCEED THE ORIGINAL PURCHASE PRICE OF THE PUMP AND SHALL NOT EXTEND TO ANY CONSEQUENTIAL LOSS OR DAMAGE.
OPERATING INSTRUCTIONS
INSERTION OF THE FLOCARE
®
INFINITY
PUMP SET
Connect the set to the feed container. Follow the instructions for use of the Flocare Infinity giving set, for setting up the enteral feeding system.
Remove the dust cap from end of pump set.
The Flocare
®
Infinity
+ pump sets are equipped with an “automatic free-flow protection”, as a consequence the set has no roller clamp.
Fill the pump set completely with feed by gently pinching on the side of the cassette that is marked with a
drop (point A on figure 5a) or fill the pump set with help of the pump (see paragraph: filling the pump set).
Open the pump door by pressing up on the lower wall and rotating the door upwards at the same time (see figure 5b).
Position the looped section of the silicone tubing around the rotor.
Stretching lightly, (see figure 5c) seat the cassette into the pump (see figure 5d).
Close the pump door
SWITCHING “ON”
For proper pump operation, make sure the pump door is closed prior to switching the
pump on by holding the “ON/OFF” key for 2 seconds otherwise the pump will generate an error.
The pump beeps and carries out a short self test, showing the pump serial number in 8 digits. Verify that all display
segments and symbols are active as shown in figure 6 (see next page).
The pump displays the total volume delivered since the memory was cleared and switches in the hold mode ready
for programming.