-21-
• Mixed or reconstituted nutrition may have foam. If using this type of nutrition, allow it to sit for some time until the foam has been dissolved before
pouring it into a feeding reservoir. This will reduce the chance of an ‘AIR in giving set’ alarm.
• Non-liquid food ingredients and mixed or reconstituted foods may tend to occlude the flow within the enteral feeding lines. If patients are advised by
their healthcare professional to use a mixture of feeds, the feed must be given when there is direct supervision (e.g. parents) to ascertain the feeding
therapy is running normally and to ensure proper actions are taken in case there is a pump failure or unexpected behaviours caused by the mix.
Technical & EMC
• The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years.
• Do not use the pump in strong magnetic fields. e.g. MRI (Magnetic Resonance Imaging) environments or areas where there is a risk of explosions
e.g. in the presence of flammable anaesthetics.
• Do not modify this equipment. Modification of this device or its accessories can result in physical hazards including delayed therapy, over delivery,
under delivery, electrocution and fire. These hazards could result in patient injury or death.
• Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC
information provided in the accompanying documents.
• Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the Flocare Infinity pump, including cables specified by the manufacturer. Otherwise, degradation of the per-
formance of this equipment could result. Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because
it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
• The Flocare Infinity pump delivers the dose within +/-5% volumetric accuracy under the following conditions:
- utilizing Flocare Infinity disposable giving sets
- fluid head height at +152 mm ± 76 mm with respect to the center of rotor.
PREPARATION
• Before using or setting a feeding regimen, check the integrity of the pump. Do not use the pump if it is damaged. If a technical failure occurs or if
the pump is dropped, the pump should be checked by a qualified technician.
• Check the position of the feeding tube, as advised by your healthcare professional, before commencing tube feeding.
• For bedside use, the multi-position pole clamp (figure 3) can be attached to the pump with the screw provided. The pump can be fixed in any
position (rotatable in 360°).
SERVICE AND WARRANTY
Service Life - Under normal conditions of use, including proper cleaning and inspection, if servicing recommendations and protocols are adhered to,
the pump has an expected lifespan of minimum 5 years. The expected service life of the battery is 2 to 5 years, depending on usage and number of
charging cycles.
The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years as indicated on the pump (indicator 9). Only
authorised personnel should perform service work on Infinity
™
+ pumps. Please contact your local Sales Organisation / Nutricia Subsidiary for all service
and repair of pumps (see address at the back of the booklet).