TECHNICAL SPECIFICATIONS
This Flocare
®
Infinity
+ pump in combination with the charger and the Flocare
®
Infinity
Nurse Call are designed to EN 60601-1-2,
EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions and immunity and are in compliance with
Directive 93/42/EEC.
0344
IEC 601: Class II Medical Device Directive: Class IIa
Microprocessor controlled
Type BF Shock Protection (EN60601-1) degree of protection against electrical shock. No electrical connection to the patient.
Drop from any height of 90cm shall not damage pump operation.
Charger: input 100-240V AC / 50-60 Hz / 0,4A Max.
output 5 V DC 3 A
Battery: internal rechargeable Lithium ion battery, 3,7 V DC 2500 mAh
Battery capacity: 24 hours at 125 ml/h for a new battery
IPX5 (EN60529 degree of protection against water entering the enclosure). Water jets from any direction shall have no effect on
the pump.
Dimensions: 140 x 95 x 35 mm
Weight: approx. 392 g
Accuracy flow rate: ±5,0% with appropriate Flocare
®
Infinity
pump set
Humidity: Operation mode: 30% to 75% noncondensing
Storage: 10% to 95 % noncondensing
Temperature: Operation mode: +5°C to +40°C
Storage and transportation: -20°C to +65°C
Atmospheric pressure: Operation mode: 70-106 kPa
Storage and transportation: 50-106 kPa
Occlusion detection pressure: Upstream occlusion: -34 kPa (tolerance 21 kPa)
Downstream occlusion: 83 kPa (tolerane 21 kPa)
Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies from 0.5 - 1.5 ml
which relates to an air bubble with an approximate length in the tubing of 9 - 26 cm.
The Flocare
®
Infinity
+ pump can safely be operated on commercial aircraft.
The use of other accessories, chargers and cables than listed within this manual may result in increased emissions or decreased
immunity of the equipment of the Flocare
®
Infinity
+ pump.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information provided in the accompanying documents.
Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equipment. If the Flocare
®
Infinity
+ pump is used adjacent to or stacked with other equipment, the pump should be observed to verify normal operation.
In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number.
At the end of their service lifetime, the pump and its electrical accessories (AC Adapter Charger and Data Download Accessory
Cable), should be disposed of according to local standards and regulations governing the disposal of electronic waste (e-waste).
The pump contains a rechargeable lithium-ion battery. When disposing of the pump, be sure to discard this equipment in a manner
consistent with local policy for expired battery operated equipment. Other accessories can be disposed of or recycled as standard
non-hazardous waste.