-34-
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation appli-
cable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
0,01
0.12 0.12 0.23
0,1
0.38 0.38 0.73
1
1.2 1.2 2.3
10
3.8 3.8 7.3
100
12 12 23
The Flocare
®
Infinity
™
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the Flocare
®
Infinity
™
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Flocare
®
Infinity
™
as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum output power of
transmitter
(W)
Recommended separation distances between portable and mobile RF communications equipment and the
Flocare
®
Infinity
™
enteral feeding pump
150 KHz to 80 MHz
d = 1.2
CD
P
80 MHz to 800 MHz
d = 1.2
CD
P
800 MHz to 6 GHz
d = 2.3
CD
P
Immunity test IEC 60601 test level Electromagnetic environment – guidance
Guidance and manufacturer’s Declaration – Electromagnetic immunity
Compliance level
The Flocare
®
Infinity
™
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare
®
Infinity
™
enteral feeding pump should assure that it is used in such an environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m Power frequency magnetic fields should be at levels characte-
ristic of a typical location in the typical commercial or hospital
environment.
30 A/m
Separation distance according to frequency of transmitter (m)