TO CONNECT WITH NURSE CALL OR PDMS
Hold the ‘+’ key down and press “ON/OFF” for 2 seconds. The pump enters the set up mode. Press the
“FILL SET” key to switch to the next setting:
“OUTP PDMS” (+) to connect with a Patient Data Monitoring System. “OUTP NRSE” to connect with a NURSE
CALL system. Use the “+” or “-” key to change to setting from “OUTP PDMS”, to “OUTP NRSE” or “OUTP
OFF” and vice versa.
Press the key and keep it pressed for 2 seconds to exit the SET UP mode and turn off the pump. The
settings are automatically saved.
TECHNICAL SPECIFICATIONS
•
This Flocare
®
Infinity
™
+ pump in combination with the adapter and the Flocare
®
Infinity
™
Nurse Call are desig-
ned to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emis-
sions and immunity and are in compliance with Directive 93/42/EEC.
0344
•
IEC 601: Class II Medical Device Directive: Class IIa
•
Microprocessor controlled
•
BF Equipment
• Charger: input 100-240V AC / 50-60 Hz / 0,4A Max.
output 5 V DC 2,4 A
•
Battery: internal rechargeable Lithium ion battery, 3,6 V DC 2000 mAh
• Battery capacity: 24 hours at 125 ml/h
•
Jet waterproof: IPX 5
•
IPX 5: Splash water proof: Water jets from any direction shall have no effect on the pump.
•
Dimensions: 140 x 95 x 35 mm
• Weight: approx. 392 g
•
Accuracy flow rate: ±5,0% with appropriate Flocare
®
Infinity
™
pump set
• Humidity: Operation mode: 30% to 75% noncondensing
Storage: 10% to 95 % noncondensing
• Temperature: Operation mode: +5°C to +40°C
Storage and transportation: -20°C to +65°C
• Atmospheric pressure: Operation mode: 70-106 kPa
Storage and transportation: 50-106 kPa
• Occlusion detection pressure: Upstream occlusion: -34 kPa (tolerance 21 kPa)
Downstream occlusion: 83 kPa (tolerane 21 kPa)
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies
from 0.5 - 1.5 ml which relates to an air bubble with an approximate length in the tubing of 9 - 26 cm.
OUTP NRSE
–
OUTP PMDS
+
fill set
2 sec
“LITE ON” (+) the light of the LCD remains on when the pump is connected to the mains. “LITE OFF” (-) the
light of the LCD switches off after a few seconds even if connected to the mains. Use the “+” or “-” key to change
the setting from “LITE ON” to “LITE OFF” or vice versa.