Flocare Pump Instruction Book & Guide 12
REASON
Air Bubble in the giving set.
The bag of feed is empty.
The air sensor is dirty.
The giving set has been used
for 24 hours or more.
The end of the giving set has
not fully pierced the foil.
The pump has administered
the dose.
SOLUTION
Disconnect the giving set from the feeding tube
and hold over a clean container. Press the FILL
SET button until the air bubble has passed out of
the giving set. Press FILL SET to stop, reattach to
feeding tube and press START.
If further feeding is required, replace the empty
pack/bottle of feed and recommence feeding. If
you need to refill the giving set with feed, follow
solution instructions for ‘‘air bubble in the giving
set’’. If required dose delivered, switch o pump
and discard feed and giving set as per training.
Clean the sensor area following the cleaning
guidelines.
Replace the giving set with a fresh one.
Press and hold the ON/OFF button until it
beeps and is powered o
If the END OF DOSE alarms inappropriately
contact Nutricia Homeward to speak to a nurse
(numbers on rear of book)
AIR
END OF
DOSE