Flocare Pump Instruction Book & Guide 8
TROUBLESHOOTING & PUMP ALARMS STEP BY STEP
No Power Supply
Light on
REASON SOLUTION
Loose wire or the plug adaptor
pins have moved.
Plug Symbol - visible while
the pump is connected to
the electrical supply
Power indicator light
Light o Light on
Check that the part where the three pins are has not
become loose and separated from the main body of the
adaptor. If this has happened, simply push back together so
the top of the charger is totally flat (see above).

Other manuals for Nutricia Flocare Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nutricia Flocare Infinity and is the answer not in the manual?

Nutricia Flocare Infinity Specifications

General IconGeneral
Device TypeEnteral Feeding Pump
ManufacturerNutricia
ModelFlocare Infinity
Accuracy±5%
Flow Rate Increment1 ml/h
Keep Vein Open (KVO)1 ml/h
Power SupplyRechargeable battery or mains power
Bolus Dose1 to 400 ml
Bolus Interval1 to 24 hours
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Empty, Low Battery, Error
Storage Temperature-20 to 60°C
Battery Life24 hours
Operating Temperature5 to 40°C