-72-
recharge the batter or select a lower rate to complete the feed ccle.
In case of batter failure, caused b, for example, an excessive temperature condition (which ma resolve
itself with time) or b a complete batter failure, the E, F and batter smbol will flash. Turn the pump “off and
contact our healthcare provider or refer the pump to service.
CONNECTION TO AN EXTERNAL ALARM SySTEM
The power connector at the side of the pump offers the option of connecting the pump to an external alarm
sstem or Patient Data Monitoring Sstem (PDMS).
The Flocare
®
Infinit
Nurse Call (corporate code 35752) enables connection to an external alarm sstem.
This external alarm sstem ma be with open or closed contact. Follow the instructions for use of the
Flocare
®
Infinit
Nurse Call to connect the pump.
The Flocare
®
Infinit
PDMS Cable (corporate code 35776) enables connection to an external PDMS
sstem. Follow the instructions for use of the Flocare
®
Infinit
PDMS Cable to connect the pump.
ALARM FUNCTIONS AND SAFETy FEATURES
In case an of the problems listed on the problem solver chart occur, the pump delivers an audible and visual
alarm and stops working. The back light of the LCD automaticall switches on.
Exception to this is the low batter alarm “BATT”, in this situation the pump continues working.
Action in the event of an alarm:
Check the tpe of alarm displaed b the large displa characters.
Press the “START/STOP ke to stop the audible and visual alarm.
Correct the cause of the alarm as described in the table.
Start the pump again b pressing “START/STOP.
The problem solver chart on the next page gives a clear explanation of the alarms.
MAINTENANCE
Cleaning:
Alwas unplug the pump prior to cleaning to avoid electric shock hazard.
On a regular basis thoroughl clean all surfaces of the pump (including the sensors and rotor) with warm
soap water, a 5% bleach solution in water, or a multipurpose disinfectant cleaner.
The Flocare
®
Infinit
pump ma be rinsed b holding under a stream of warm, clean water. Do not sub-
merge the pump!
Alwas maintain the rollers on the rotor in a clean state to ensure the spin freel.
The adaptor normall does not require cleaning. When desired, a dr or slightl damp cloth ma be used
to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.
Safet notes:
The manufacturer recommends an inspection of the pump at an authorised service centre ever 2 ears.
If an fault occurs during use, or if the pump is dropped, it should be checked b authorised technical per-
sonnel prior to use.
Replace the Flocare
®
Infinit
pump set ever 24 hours to maintain deliver accurac and prevent the
growth of harmful bacteria. Dispose of Flocare
®
Infinit
disposable sets properl, as required b local law.
Do not use the Infinit alarm sstems to trigger actions related to secondar (electrical) medical devices
(e.g. a volumetric or sringe pump).
Do not use pump functions (e.g. the fill set function) for an other purpose than described in this manual,
as this ma cause the pump to incorrectl calibrate.