Nutricia Flocare Infinity - Changing Feeding Program; Clearing Pump Memory; Battery Operation and Charging

Nutricia Flocare Infinity
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING
Press the “START/STOP” ke to pause the pump.
Change the program b pressing the required ke (ml/h or DOSE=VOL) and adust using the “+” or “-” kes.
Restart the pump b pressing the “START/STOP” ke again.
CLEARING THE MEMORy
All parameters and values can be cleared one after the other in the following wa:
Press “START/STOP” to pause the pump.
Select the parameter that needs to be cleared.
Press the CLR” ke to clear the memor. The parameter or value returns to its default value:
ml/h = 0 ml/h
VOL = cont = No dose set. Pump will feed continuousl until feed container is empt or pump
is stopped
INFO = 0 ml = Volume delivered is cleared
In order to keep a clear overview of the dail amount of feed delivered, clear the total volume delivered as
each dail feeding schedule is started, as follows:
Switch the pump “ON”
Immediatel after the self-test the pump displas total volume delivered.
Press the “CLR” ke. The total volume delivered returns to “0 ml”.
The pump displas the previous programmed flow rate (ml/h).
The pump is in the “HOLD” mode read for programming.
When another feeding program needs to be started, or the pump will be used for another patient, all set-
tings (rate and dose) can be returned to default and the volume delivered can be set at zero b simpl
pressing the “CLR” ke and keeping it pressed for 2 seconds.
BATTERy OPERATION
The Flocare
®
Infinit
pump is equipped with a Lithium-ion batter.
The batter smbol in the displa is automaticall replaced b the plug smbol when the adapter is
plugged in. The “fuel gauge” then indicates that the batter is charging b displaing the segments in
an ascending low to high pattern starting with the leftmost segment.
This pattern continuousl repeats while the pump is charging.
To check the status of the batter, disconnect the charger from the pump and turn the
pump on. The bars between E and F (E=Empt, F=Full) represent the “fuel gauge” of the
batter. Each bar is approximatel 1/4 of a full batter charge. If 2 bars appear the
batter is half full and remaining operating time is approximatel 12 hours at a flow rate
of 125 ml/h.
In case of power failure the pump automaticall switches to batter powered operation.
When during functioning the last block is gone, the batter smbol will blink to indicate that there is approxima-
tel 1 hour of charge left. The displa will flash “BATT ever 3 seconds, alternating with the active displa and
the pump will beep ever 2 seconds to remind the user of the low batter charge condition. Plug in the charger
to continue to run and recharge the batter.
As the batter level decreases the pump ma not be able to deliver tube feeds at rates greater than 300 ml/h,
even though some batter capacit remains. The pump will alarm and indicate
“BATT”. If this happens, either
-71-
START/STOP
CLR
d
UK-IE
ZMNL400336_AW.indd 73 3/18/10 3:47:43 PM