PROGRAMMING THE PUMP
• The Flocare
®
Infinit
™
pump can be used for continuous or dose feeding.
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessar (see para-
graph: filling the pump set).
• The displa shows the total volume delivered since the pump was last cleared.
• If necessar clear the volume delivered b pressing the “CLR” ke.
• The last installed flow rate (ml/h) is now displaed.
• The pump is now in the “hold” mode and is read for programming.
CONTINUOUS FEEDING
• Adust the flow rate (ml/h) if required b using the “+” or “-” ke. Hold either ke down to change rapidl.
• If another parameter is indicated in the LCD, it will be necessar to first press the “ml/h” ke followed b the
“+” or “-” ke to set the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortl at 50 - 125 - 250 ml/h.
• Press the “DOSE=VOL” ke and make sure the value is put at 0 ml indicated b the word “CONT”. This
means the pump will run in a continuous mode.
• Start the pump b pressing “START/STOP”.
• The arcs around the word “run” start rotating.
• The flow rate (ml/h) is displaed whilst functioning.
TO
SET A DOSE
From the hold mode, with the last flow rate (ml/h) displaed in the LCD:
• Set the flow rate b pressing the “+” or “-” ke.
If another parameter is displaed:
• Press first the “ml/h” ke followed b “+” or “-” ke to adust the flow rate. The flow rate can be set between
1 and 400 ml/h, with increments of 1ml.
• Press the “DOSE=VOL” ke and install the volume to be administered with the “+” and “-” ke. The range
goes from 1 - 4000 ml, with steps from 1ml.
• Start the pump b pressing “START/STOP”.
During functioning the following parameters can be found in the displa:
• B pressing the “ml/h” ke the “flow rate” is displaed.
• B pressing the “INFO” ke the total volume since the memor was cleared will be
visualized.
When the required volume or dose has been delivered, “END OF DOSE” will appear and the pump will beep
(or mute, depending upon the pumps’ configuration (see paragraph: set up mode).
ON/OFF
d d d d d
ml/h
START/STOP
400 ml/h
+
1 ml/h
-
4000 ml
+
1 ml
-
DOSE=VOL
ON/OFF
d d d
ml/h
400 ml/h
+
1 ml/h
-
START/STOP
INFO