EasyManua.ls Logo

Nutricia Flocare Infinity - Filling the Pump Set; Display Information

Nutricia Flocare Infinity
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-69-
to administer medication without switching the pump off,
to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message “PUSH STRT appears in the displa.
Press “START/STOP” to stop the alarm and to extend the hold mode b a further 3 minutes or press “START/
STOP” twice (3 beeps will be heard) to resume programmed settings.
FILLING THE PUMP SET
The FILL SET function offered b the Infinit pump is used to fill a (new) feeding set with
nutrition (or water as the case ma be). When the FILL SET function runs all alarms
are deactivated helping avoid alarms being inadvertentl triggered (e.g. air alarm).
The FILL SET function should onl be used to fill an empt (air-filled) feeding set. Use
of the FILL SET function at an other time ma cause the pump to incorrectl calibrate.
When the FILL SET function is initiated the pump will calibrate to the giving set.
When the pump is in the hold mode:
Press and hold the “FILL SET” ke for 2 seconds to activate the “FILL SET” maneuver. Release the button as
soon as the pump generates a beep and starts pumping at a flow rate of approx. 700ml/h. During this maneu-
ver the message “FILL SET” appears in the displa. The pump will automaticall stop when the Flocare
®
Pack
Infinit
pump set is completel filled with feed. However this “FILL SET” maneuver can at antime be stopped
b pushing the “FILL SET” ke a second time. The pump will return to the hold mode when the “FILL SET”
maneuver is complete or stopped.
DISPLAy
The pump has a liquid crstal displa (L.C.D.) with large alphanumeric characters, smaller smbols, words
and a back light.
The following information can be found in the displa:
Flow rate (ml/h), volume (ml), is displaed through the large characters.
Words below describe what the number relates to (rate, dose or volume).
The pump also displas messages, for example “end of dose” will appear when the pump has finished
delivering a single feed dose.
The arcs around the run smbol rotate when the pump is
running.
The remaining batter capacit (see paragraph: batter
operation).
The wall plug smbol indicates that the charger is plugged in.
The back light of the displa turns off 10 seconds after the last ke is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
Figure 6
ON/OFF
FILL SET
d
UK-IE
ZMNL400336_AW.indd 71 3/18/10 3:47:42 PM

Other manuals for Nutricia Flocare Infinity

Related product manuals