OPERATING INSTRUCTIONS
INSERTION OF THE FLOCARE
®
INFINITy
PUMP SET
Connect the set to the feed container as instructed on the packaging of the Flocare
®
Infinit
pump set.
Remove the dust cap from the step connector.
The Flocare
®
Infinit
pump sets are equipped with an “automatic free-flow protection”, as a consequence
the set has no roller clamp.
Fill the pump set completel with feed b gentl pinching on the side of the cassette that is marked with a
drop (point A on figure 5a) or fill the pump set with help of the pump (see paragraph: filling the pump set).
Open the pump door b pressing up on the lower wall and rotating the door upwards at the same time
(see figure 5b).
Position the looped section of the silicone tubing around the rotor.
Stretching lightl, (see figure 5c) seat the cassette into the pump (see figure 5d).
Close the pump door.
SWITCHING “ON”
For proper pump operation, make sure the pump door is closed prior to switching the
pump on. Press the “ON/OFF” ke for 2 seconds.
The pump beeps and carries out a short self test, showing the pump serial number
in 8 digits. Verif that all displa segments and smbols are active as shown in figure
6 (see next page).
The pump displas the total volume delivered since the memor was cleared and
switches in the hold mode read for programming.
SWITCHING “OFF”
Press the “ON/OFF” ke and keep it pressed for 2 seconds. A continuous alarm will be
heard and the pump switches off.
The feeding program (installed parameters) and total volume administered since the
last clearance will be retained in the pump’s memor.
If the pump shuts down due to low batter voltage, the memor will be retained for
24 hours.
“HOLD” MODE
To temporaril pause the pump, or switch into “hold” mode whilst operating, press the
“START/STOP” ke once. Three beeps are heard and the run smbol disappears.
The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume
(ml) are retained.
The “hold” mode is used to temporaril stop the flow of feed:
to change the feeding program (installed parameters),
to change the feed container,
ON/OFF
ON/OFF
START/STOP