SET UP MODE
The set up mode is used to set the alarm level,
to lock the kepad,
to switch the audible alarm off when a dose is administered,
to switch the light of the LCD permanentl on when connected to the mains,
to enable connection with a Nurse call or Patient Data
Monitoring Sstem (PDMS).
This set up mode ma onl be entered b healthcare professionals and authorised personnel trained to use
this application of the pump.
The patient or his/her relatives without permission of the phsician, dietician, nurse or other licensed practi-
tioner, ma not change the settings of the set up mode.
TO SET THE ALARM LEVEL
Switch the pump “on” while holding the “+” ke down. The pump enters the set up mode.
First the alarm level can be set:
“BEEP HIGH” (+) or “BEEP LOW” (-)
Use the “+” or “-” ke to change the setting from “high” to “low” or from “low” to “high”.
TO
LOCK THE KEyPAD
Press in the set up mode the “FILL SET” ke to switch to the next setting:
“UNLK” the kepad (-) or “LOCK” the kepad (+). In the “LOCK” mode the feeding program of the pump
can no longer be changed.
Use the “+” or “-” ke to change the setting from “lock” to “unlk” or vice versa.
ON/OFF
&
d
Beep high
+
Beep low
-
+
2 sec 2 sec
FILL SET
d
Lock
Unlk
-
+
ER01 - ER99
No plug smbol visi-
ble, while the pump
is connected to the
mains.
Condition
LOCK
The self test detected an electronic
error
The wall outlet doesn’t work.
The adaptor is damaged.
Cause
Onl the current feeding
schedule is allowed for this patient.
Another feeding program is not
allowed b our healthcare profes-
sional.
Turn the pump “off”, make sure the pump door is closed
and switch the pump back “on”. If the error persists,
contact our healthcare provider or refer the pump to
service.
Connect the pump to another wall outlet.
Contact our healthcare professional or
Nutricia subsidiar to replace the adaptor.
Correction
The programming feature is blocked in the set up mode
of the pump. Ask our healthcare professional to modif
this setting.