TO
MUTE WHEN DOSE DONE
Press in the set up mode the “FILL SET” ke to switch to the next setting:
“BEEP WHEN DONE” (+): the pump will give an alarm when the dose is administered or “MUTE WHEN
DONE” (-): the pump will not give an alarm after administering the dose. Use the “+” or “-” ke to change
the setting from “beep when done” to “mute when done” or vice versa.
TO SWITCH LIGHT ON
Press in the set up mode the “FILL SET” ke to switch to the next setting:
“LITE ON” (+) the light of the LCD remains on when the pump is connected to the mains. “LITE OFF” (-) the
light of the LCD switches off after a few seconds even if connected to the mains. Use the “+” or “-” ke to change
the setting from “lite on” to “lite off” or vice versa.
TO
CONNECT WITH NURSE CALL OR PDMS
Press in the set up mode the “FILL SET” ke to switch to the next setting:
“OUTP PDMS” (+) to connect with a Patient Data Monitoring Sstem. “OUTP NRSE” to connect with a NURSE
CALL sstem. Use the “+” or “-” ke to change to setting from “OUTP PDMS”, to “OUTP NRSE” or “OUTP
OFF” and vice versa.
Press the “ON/OFF” ke and hold it down for 2 seconds to exit the SET UP mode. The settings are auto-
maticall saved.
TECHNICAL SPECIFICATIONS
This Flocare
®
Infinit
pump in combination with the adapter and the Flocare
®
Infinit
Nurse Call are designed
to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions
and immunit and are in compliance with Directive 93/42/EEC.
0344
IEC 601: Class II Medical Device Directive: Class IIa
Microprocessor controlled
BF Equipment
Charger: input 100-240V AC / 50-60 Hz / 0.4A Max.
output 5 V DC 2.4 A
Batter: internal rechargeable Lithium ion batter, 3.6 V DC 2000 mAh
Batter capacit: 24 hours at 125 ml/h
jet waterproof: IPX 5
IPX 5: Splash water proof: Water ets from an direction shall have no effect on the pump.
-75-
FILL SET
d
Beep when done
Mute when done
-
+
FILL SET
d d
OUTP PDMS
OUTP NRSE
-
+
ON/OFF
2 sec
FILL SET
d d
Lite on
Lite off
-
+
ON/OFF
2 sec
UK-IE
ZMNL400336_AW.indd 77 3/18/10 3:47:43 PM