-76-
•
Dimensions: 140 x 95 x 35 mm
• Weight: approx. 392 g
• Accurac flow rate: ±5,0% with appropriate Flocare
®
Infinit
™
pump set
• Humidit: Operation mode: 30% to 75% noncondensing
Storage: 10% to 95 % noncondensing
• Temperature: Operation mode: +5°C to +40°C
Storage and transportation: -20°C to +65°C
• Atmospheric pressure: Operation mode: 70-106 kPa
Storage and transportation: 50-106 kPa
• Occlusion detection pressure: Upstream occlusion: -34 kPa (tolerance 21 kPa)
Downstream occlusion: 83 kPa (tolerane 21 kPa)
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies
from 0.5 to 1 ml which relates to an air bubble with an approximate length in the silicone pump segment of
6 - 13 cm.
• The Flocare
®
Infinit
™
can safel be operated on commercial aircraft.
• The use of other accessories, adaptors and cables than listed within this manual ma result in increased
emissions or decreased immunit of the equipment of the Flocare
®
Infinit
™
pump.
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in the accompaning documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equip-
ment. If the Flocare
®
Infinit
™
pump is used adacent to or stacked with other equipment, the pump should
be observed to verif normal operation.
• In case of pump scrapping, alwas notif our Nutricia sales unit of the pumps serial number. Pump scrap-
ping should alwas occur according to local legislation.