Emissions test Compliance Electromagnetic environment – guidance
Guidance and manufacturer’s Declaration – Electromagnetic emissions
Immunit test IEC 60601 test level Electromagnetic environment – guidance
Guidance and manufacturer’s Declaration – Electromagnetic immunit
Compliance level
161 V
ac
for 0,5 s
95 V
ac
for 100 ms
0 V
ac
for 10 ms
0 V
ac
for 5 s
161 V
ac
for 0,5 s
95 V
ac
for 100 ms
0 V
ac
for 10 ms
0 V
ac
for 5 s
APPENDIX A: GUIDANCE AND MANUFACTURER’S DECLARATION -
ELECTROMAGNETIC COMPATIBILITy
The Flocare
®
Infinit
™
enteral feeding pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such an
environment.
RF Emissions
CISPR 11
Group 1 The Flocare
®
Infinit
™
uses RF energ onl for its internal
function. Therefore, its RF emissions are ver low and are
not likel to cause an interference in nearb electronic
equipment.
RF Emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
The Flocare
®
Infinit
™
is suitable for use in all
establishments, including domestic establishments and
those directl connected to the public low-voltage power
suppl network that supplies buildings used for domestic
purposes.
The Flocare
®
Infinit
™
enteral feeding pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the Flocare
®
Infinit
™
enteral feeding pump should assure that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
≤ 6 KV contact
≤ 8 KV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with snthetic material, the
relative humidit should be at least 30%.
≤ 6 KV contact
≤ 8 KV air
Electric fast
transient/burst
IEC 61000-4-4
≤ 2 KV for power sup-
pl lines
≤ 1 KV for input/ out-
put lines
Mains power qualit should be that of a tpical
commercial or hospital environment.
≤ 2 KV for power sup-
pl lines
≤ 1 KV for input/ out-
put lines
Surge
IEC 61000-4-5
≤ 1 KV differential
mode
Mains power qualit should be that of a tpical
commercial or hospital environment.
≤ 1 KV differential
mode
Voltage dips, short
interruptions and
voltage variations on
power suppl lines.
IEC 61000-4-6
Mains power qualit should be that of a tpical
commercial or hospital environment. If the user
of the Flocare
®
Infinit
™
requires continued
operation during power mains interruptions, it is
recommended that the Flocare
®
Infinit
™
enteral
feeding pump be powered from an uninterrupti-
ble power suppl of batter.