-79-
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequenc of the transmitter, where P is the maximum output
power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequenc range applies.
NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and
reflection from structures, obects and people.
0,01
0,01 0,01 0,02
0,1
0,04 0,04 0,07
1
0,12 0,12 0,23
10
0,37 0,37 0,74
100
1,17 1,17 2,33
The Flocare
®
Infinit
™
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF dis-
turbances are controlled. The customer or the user of the Flocare
®
Infinit
™
can help prevent electromagnetic interference
b maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Flocare
®
Infinit
™
as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum output
power of transmitter
(W)
Separation distance according to frequenc of transmitter (m)
Recommended separation distances between portable and mobile RF communications euipment and the
Flocare
®
Infinit
™
enteral feeding pump
150 KHz to 80 MHz
d = 0,12
CD
P
80 MHz to 800 MHz
d = 0,12
CD
P
800 MHz to 2,5 GHz
d = 0,23
CD
P
Portable and mobile RF communications equi-
pment should be used no closer to an part of
the Flocare
®
Infinit
™
, including cables, than the
recommended separation distance calculated
from the equation applicable toe the frequenc
of the transmitter:
Recommended separation distance
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined b an electromagnetic site surve,
a
should be less than the compliance level in
each frequenc range.
b
Interference ma occur
in the vicinit of equipment
marked with the following
smbol:
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80 MHz
d = 1,17CD
P
3 Vrms
150 KHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 Mhz to 2,5 GHz
d = 0,12CD
P
d = 0,23CD
P
80 MHz to 800 MHz
800 MHz to 2,5 GHz
30 V/m
80 Mhz to 2,5 GHz
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoreticall with accurac.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site surve should be
considered. If the measurement field strength in the location in which the Flocare
®
Infinit
™
is used exceeds the appli-
cable RF compliance above, the Flocare
®
Infinit
™
should be observed to verif normal operation. If abnormal perfor-
mance is observed, additional measures ma be necessar, such as reorienting or relocating the Flocare
®
Infinit
™
.
b. Over the frequenc range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
NOTE 1 At 80 Mhz and 800 MHz, the higher frequenc range applies.
NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and
reflection from structures, obects and people.
The Flocare
®
Infinit
™
enteral feeding pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the Flocare
®
Infinit
™
enteral feeding pump should assure that it is used in such an environment.
Guidance and manufacturer’s Declaration – Electromagnetic immunit
Immunit test IEC 60601 test level Electromagnetic environment – guidanceCompliance level
ZMNL400336_AW.indd 81 3/18/10 3:47:44 PM