5 Cleaning and Disinfection
6 / 26 Conditioning Instruction BIOM® 5 (G/55185/0818/en)
5 Cleaning and Disinfection
Recommendation:
Clean or disinfect the BIOM® 5 by machine, as this is much more
effective.
Start pre-conditioning immediately after receiving the equipment in
the conditioning station.
5.1 Preparation: Dismantle the BIOM
®
5
You must dismount the BIOM® 5 before you condition it.
Pull the BIOM® 5 off the adaptor plate. To do so,
swing out the BIOM® 5
Take out the reduction lens and put it down onto
the soft pad.
Pull off the front loupe and put it down onto the
soft pad.
Fig. 5–1: Dismounting the components
1
2
3