10 Overview: Permitted Sterilisation Method – List of Articles
22 / 26 Conditioning Instruction BIOM® 5 (G/55185/0818/en)
10 Overview: Permitted Sterilisation Method – List of Articles
Article Name Article
No.
Cleaning
Method
Approved Sterilisation Processes
Steam
Autoclave
STATIM
5000S/G4
Plasma/
STERRAD
®
100S SHORT
cycle
Plasma/
STERRAD
®
100NX
®
Standard
cycle
V-Pro® Low
Temperature
Sterilization
System
1
134° C
(273.2° F)
132° C
(269.6° F)
134° C (273.2° F);
3,5 min
Sterilisation
container
55180
manual,
by machine
x x
Insert for sterili-
sation container
55185
manual,
by machine
x x x x x
Insert for sterili-
sation container
55186 x x x x
BIOM® 5m 55462
manual,
by machine
x x x x x x
BIOM® 5ml 55463
manual,
by machine
x x x x x x
BIOM® 5c 55400
manual,
by machine
x x x x x x
BIOM® 5cl 55403
manual,
by machine
x x x x x x
Antriebsriemen
BIOM® 5c/cl
54176
manual,
by machine
x x x x x x
Cable conduct
BIOM® 5
54178
manual,
by machine
x x x max.
10 times
max.
10 times
x
Adapter
Adapter plate 55423
manual,
by machine
x x x x x x
Adapter plate 55424
manual,
by machine
x x x x x x
Adapter plate 55426
manual,
by machine
x x x x x x
Adapter plate 55431
manual,
by machine
x x x x x x
Adapter plate 55425
manual,
by machine
x x x x x x