5 Cleaning and Disinfection
Conditioning Instruction BIOM® 5 (G/55185/0818/en) 13 / 26
5.4 Manual Cleaning/Disinfection
The method described below has been validated for manual cleaning/
disinfection:
If you clean and disinfect your equipment manually, you must verify your
procedure independently and validate it specific to the product and the
method.
Cleaning tools
Cleaning solution with 0.8% detergent (Cidezyme/Enzol from
Johnson & Johnson)
Water
Procedure
Perform a manual disinfection in accordance with the requirements
of the VAH or the applicable standards.
Check that the components have no visible signs of soiling, sect. 5.2,
page 8.
Place the components in the cleaning solution for 5 minutes at 40°C.
The BIOM® 5 must be fully covered.
Rinse the components for 5 seconds under running water (static
pressure 4.2 bar).