5 Cleaning and Disinfection
12 / 26 Conditioning Instruction BIOM® 5 (G/55185/0818/en)
OCULUS Sterilisation Insert
You can use the specially designed sterilisation insert, OCULUS Art.
No. 55185 for the BIOM® 5 and its accessories.
One adaptor plate, the BIOM® 5, one reduction lens, at least one front
loupe, 2 rubber drive belts, sterile cap for the knurled screw, SDI® control
knob belong together. These components can be put into the special
insert and placed into a suitable sterilisation container for sterilisation in
the steam autoclave.
Fig. 5–4: Components of the BIOM® 5 on the sterilisation insert
Dimensions of the OCULUS sterilisation insert
Width 240 mm
Depth 245 mm
Height 53 mm
Total height, loaded 80 mm