10 Overview: Permitted Sterilisation Method – List of Articles
Conditioning Instruction BIOM® 5 (G/55185/0818/en) 23 / 26
Article Name Article
No.
Cleaning
Method
Approved Sterilisation Processes
Steam
Autoclave
STATIM
5000S/G4
Plasma/
STERRAD
®
100S SHORT
cycle
Plasma/
STERRAD
®
100NX
®
Standard
cycle
V-Pro® Low
Temperature
Sterilization
System
1
134° C
(273.2° F)
132° C
(269.6° F)
134° C (273.2° F);
3,5 min
Adaptor Accessories
Sterile cap for
knurled screw
54580 manual x x
x x x x
SDI
SDI® II e
SDI® II m
SDI® 3 c
SDI® 4m
SDI® 4e
SDI® 4c
54800
54802
54805
54810
54812
54815
54820
54830
54300
54302
54305
54310
54312
54315
54320
54330
54331
54332
Sterilisation
not
possible
Wipe
disinfection
permitted,
as for
surgical
microscope
Rubber cap for SDI®
II/3 star knob
05090090
manual,
by machine
x x
x x
Rubber cap for SDI®
4 star knob
54335
manual,
by machine
x x
x x