10 Overview: Permitted Sterilisation Method – List of Articles
24 / 26 Conditioning Instruction BIOM® 5 (G/55185/0818/en)
1
tested V-Pro® systems (sterilizing agent VAPROX® HC):
Article Name Article
No.
Cleaning
Method
Approved Sterilisation Processes
Steam
Autoclave
STATIM
5000S/G4
Plasma/
STERRAD
®
100S SHORT
cycle
Plasma/
STERRAD
®
100NX
®
Standard
cycle
V-Pro® Low
Temperature
Sterilization
System
1
134° C
(273.2° F)
132° C
(269.6° F)
134° C (273.2° F);
3,5 min
Reduction Lenses
Reduction lens for
f = 175 mm on
BIOM® 5m/c
55401
manual,
by machine
x x x x
x x x
Reduction lens for
f=200 on
BIOM® 5ml/cl
55404
manual,
by machine
x x x x
x x x
Reduction lens for
f=200 on
BIOM® 5m/c
55405
manual,
by machine
x x x x
x x x
Front Lenses for BIOM
®
5
Hi-res macula lens 53606
manual,
by machine
x x x x
x x x
WiFi-HD mini lens 53605
manual,
by machine
x x x x
x x x
90 D lens 53604
manual,
by machine
x x x x
x x x
WiFi-HD lens 53603
manual,
by machine
x x x x
x x x
Wide-field lens 53602
manual,
by machine
x x x x
x x x
Mini Wide-Field lens
53601
manual,
by machine
x x x x
x x x
V-Pro® 1 Low Temperature Sterilization System
V-Pro® 1 Plus Low Temperature Sterilization System
V-Pro® 1 maX Low Temperature Sterilization System