Diagnostics
Generator Defaults
After normal bootup, the C-Arm defaults to the following conditions:
Alarm Reset is OFF.
Automatic contrast and brightness are ON.
Auto technique is ON.
Camera iris is OPEN.
Collimator leaves are OPEN.
The Collimator is not rotated.
Film mode is OFF.
Low Dose is OFF.
Magnification is NORMAL.
Camera rotation is set to customer specification.
Invert top to bottom is OFF.
Reverse left to right is OFF.
Pulse mode is OFF.
C-Arm Control Panel VFD Messages
There are five kinds of messages that appear on the C-Arm Control Panel Vacuum Fluorescent Displays:
1. A
Timed
message appears for a set time and then disappears.
2. An
Alternating
message appears for two seconds, followed by the technique, which appears for two seconds. The pattern
repeats.
3. A
User Input Required
message won’t leave the C-Arm Control Panel display until you enter the information being requested.
4. A
Disables X-Rays
message prevents the generator from producing X-rays.
5. A
System Shutdown
message appears when software prevents the generator from working.