Diagnostics
Generator Defaults
After normal bootup, the C-Arm defaults to the following conditions:
• Alarm Reset is OFF.
• Automatic contrast and brightness are ON.
• Auto technique is ON.
• Camera iris is OPEN.
• Collimator leaves are OPEN.
• The Collimator is not rotated.
• Film mode is OFF.
• Low Dose is OFF.
• Magnification is NORMAL.
• Camera rotation is set to customer specification.
• Invert top to bottom is OFF.
• Reverse left to right is OFF.
• Pulse mode is OFF.
C-Arm Control Panel VFD Messages
There are five kinds of messages that appear on the C-Arm Control Panel Vacuum Fluorescent Displays:
1. A
Timed
message appears for a set time and then disappears.
2. An
Alternating
message appears for two seconds, followed by the technique, which appears for two seconds. The pattern
repeats.
3. A
User Input Required
message won’t leave the C-Arm Control Panel display until you enter the information being requested.
4. A
Disables X-Rays
message prevents the generator from producing X-rays.
5. A
System Shutdown
message appears when software prevents the generator from working.