Diagnostics
any X-ray in progress, and disables future X-rays. First, turn the system off for at least five seconds and restart it. If the problem
persists, check the functionality of the system and check the ARCNET communications. Frequent Communication Failed errors
may also be indicative of corrupted persistent object files, which may require a software reload.
CONTROL PANEL ERROR
- The system generates this message when it cannot communicate with the C-Arm Control Panels.
The message never appears on the C-Arm Control Panel displays, but the system logs it in the Generator Error log. This error
causes the system to immediately shut down. First, turn the system off for at least five seconds and restart it. Next, check the
functionality of the system and check the ARCNET.
CONTROL PANEL KEY STUCK - RELEASE, THEN PRESS ANY KEY TO CONTINUE
- The system generates this message
when a C-Arm Control Panel key is stuck on at bootup. This disables all generator functions, including X-rays, until you correct
the sticky problem key and clear the message.
DATA ERROR
- This message appears when the system detects a data error. This can be an invalid write to an ABS table, an
invalid read from an ABS table, corrupted data in non-volatile memory, a software problem, etc. Data errors shut the system
down.
FAST STOP ACTIVATED
POWER OFF. WAIT 5 SECONDS. -
This message appears when the Interlocks are open. This is
typically the result of pressing either
FAST STOP
switch on the front of the C-Arm Control Panel. This condition disables X-ray
generation and most system functions, including column lift motors. You must power the system down, wait 5 seconds and reboot
to resume operation. If the problem persists, the 24V interlock loop has most likely been broken.
FILAMENT CAL REQUIRED - PRESS ANY KEY
- This message appears at system startup if the software detects that a
filament calibration has not been performed on the system. It is possible to receive multiple mA errors as a result of this
condition.. This error disables the generator. Press any C-Arm Control Panel key to re-enable the generator and clear the
message. If the message persists, use RUT to perform a filament calibration.
FILAMENT REGULATOR FAILURE - PRESS ANY KEY
- This message appears when the system senses that the Xray tube
filament current is out of tolerance. You may press any C-Arm control panel key to continue. However, the resolution of future
images may be degraded. If problem persists, check filament regulator circuitry.
FILAMENT SELECT ERROR - PRESS ANY KEY
- This message appears when the filament size selected in the software does
not match the current filament in the PIO hardware. Press any C-Arm control panel key to continue. X-rays are disabled while this
error is active. If problem persists, check filament select circuitry.
FILAMENT STANDBY
FLUORO ALARM
9