Calibration
Collimator / Camera Calibration
Note: For pre-OPIS system software, use the RUT utility CD.
Using RUT
1. Attach the beam alignment tool to the image tube.
2. Connect the RUT cable to the laptop and the back of the workstation.
3. Run the RUT program.
4. From the Tools pull down menu select Connect, then Upload and Save to save all data from the 9800 to the laptop.
5. Select the Collimator / Camera form
6. Click on the Calibrate Collimator Stops button and follow the instructions that appear on the laptop.
7. Click on the Normal Mode button and follow the instructions that appear on the laptop.
8. Click on the Mag Modes button and follow the instructions that appear on the laptop.
9. Click on the Calibrate Camera Stops button and follow the instructions that appear on the laptop.
10. Click on the Enable Camera Adjustment button and follow the instructions that appear on the laptop.
11. From the File pull down menu select the Save Collimator / Camera Calibration File to save the new calibration data.
12. From the File pull down menu select the Download Collimator / Camera Data to 9800 to apply the data to the 9800.
13. From the Tools pull down menu, select Disconnect.
14. Turn the 9800 system off, wait for about 10 seconds then turn the system back on. If collimator related error messages
appear, the data downloaded was bad or the download was unsuccessful.
15. Repeat the Collimator Alignment Verification in the preceding section.