Replacement
You must remove generator covers in the following sequence:
1. Front cover
2. Right side cover
3. Rear cover
4. Left side cover
5. Front leg cover
You can remove the Cross Arm Bearing cover, X-ray Tube connector cover, and CCD Camera
cover in any order. You cannot
remove the Collimator cover until the X-ray Tube Connector cover is off. You must remove the Cross Arm Bearing cover before
you can remove either Control Panel cover.
Front Cover
WARNING: Removing the front cover exposes dangerous AC line voltages on the Power/Motor
Relay PCB. Disconnect Workstation from AC line before touching this PCB.
You must remove the front cover before you can remove the right side cover, rear cover, left side cover, or front leg cover.
Proceed as follows:
1. Disconnect Workstation’s AC power plug from AC outlet.
2. Loosen four screws that secure front cover to C-Arm. Screws are located in horizontal top portion of front cover under a
flexible collar that surrounds vertical column.
3. Lift cover away from C-Arm.
5
Service
Periodic Maintenance
Contents
Schematics
Illustrated Parts
Installation