Replacement
WARNING Heat sinks on PS1 are at AC line potential.
WARNING: Dangerous AC line voltage is present on J1/P1.
Follow these steps to replace PS1:
1. Turn off Workstation and disconnect Workstation’s AC power plug from AC outlet.
2. Remove Left Side Cover.
3. Remove and retain two torx screws that secure swing-out panel to left-hand side of C-Arm chassis.
4. Disconnect P1 from J1.
5. Disconnect P2 from J2.
6. Remove torx screw and washer that secure cable guide on side of PS1. Retain screw, washer, and cable guide for reuse.
7. Remove and retain four torx screws and lockwashers that secure PS1 to swing-out panel.
8. Lift PS1 out of C-Arm.
9. Inspect replacement power supply assembly for any obvious damage or missing components.
10. Mount replacement power supply assembly to chassis swing-out panel and secure mounting hardware.
11. Connect P1 to J1 and P2 to J2 on replacement power supply.
12. Mount cable guide removed in step 6 to replacement power supply. Secure screw and lockwasher.
13. Connect Workstation’s AC power plug to AC outlet.
14. Check system for proper operation.
15. Close swing-out panel and secure with torx screws and washers removed in step 3.
16. Reinstall Left Side Cover.