Replacement
X-ray On Pushbutton
The X-ray On pushbutton is part of a replaceable cable assembly that mounts in the Cross Arm Bearing cover as shown next.
The following procedure describes how to replace the X-ray On Pushbutton.assembly.
1. Turn off Workstation and remove its power plug from facility's AC outlet.
2. Remove Cross Arm Bearing Cover as described elsewhere in this chapter.
3. Loosen suspect switch by removing its locking ring and pushing the switch upward through Cross Arm Bearing cover.
4. Disconnect switch cable and tag connector.
5. Install replacement switch and connect its cable to connector you tagged in step 4.
6. Install Cross Arm Bearing cover.
7. Power up system and test replacement switch for proper operation.