EasyManua.ls Logo

OEC UroView 2800 - Generator LEDs;Lamps Filament Heater Boards (2)

Default Icon
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OEC UroView
®
2800 Table/Generator Functional Block Diagrams
21
Service
Periodic Maintenance
Contents
Schematics
Illustrated Parts
Installation
g
Page 4 of 10
GEN_LEDS.DS4
2800
GE OEC Training
Generator LEDS/Lamps
Filament Heater Boards (2)
J2
LF
COMMON
SF
J3
J1
GND2
DS3
DS4
ON
CF2
CF1
GND5
GND1
GND6
GND4
GND3
CF3
CF4
SF_LF
RESET
DS2
DS1
LEDs:
DS1-YELLOW-STATUS-BLINKING
DS2-YELLOW-STATUS-BLINKING
DS3-GREEN-+160VDC PRESENT
DS4-YELLOW-INVERTER OUTPUT RUNNING
ON-YELLOW-INVERTER COMMAND ON
SF_LF-YELLOW-SMALL/LARGE FOCAL SPOT RELAY FEEDBACK
RESET-RED-PCB POWER UP OR RESET
GND1-6 GROUND PADS
6
4
JUMPER (ONLY ON ONE BOARD)
TELLS SOFTWARE THIS IS THE
SECOND BOARD (SMALL SPOT)
IN A TWO BOARD SYSTEM
HEAT_160V
HEAT_0V
09/02
Generator LEDs/Lamps Filament Heater Boards (2)