7
7. Electrical problems in the wiring or the heaters may not be obvious by
looking at the equipment. A malfunction is indicated if the unit starts blow-
ing fuses or tripping breakers, the heating time seems too long, or the
thermostat control does not repeat. These conditions indicate an elec-
trical repair job may be required. Always disconnect the power cables
before attempting any repair.
8. The ltration cell assembly is a pressure vessel and these safety precau-
tions should be followed:
a. Cell bodies that show signs of stress cracking, severe pitting or have
damaged threads must not be used.
b. Cell caps showing evidence of damaged or deformed threads must
not be used.
It is strongly recommended the instruction manual be attached to the
apparatus and read completely prior to the initial operation by anyone
unfamiliar with the equipment.