CHAPTER 3 MAINTENANCE PROCEDURES
FC5515R Centrifuge Service Manual 3-7 Ohaus Corporation www.ohaus.com
3.3.3 Replacing the Display
Loosen all electrical contacts to the display on the back of the front housing.
Remove the 4 fastening screws of the display and pull the display upwards
3.3.4 Replacing the foil Key Board
Remove the incremental shaft encoder and the display. Peel off the defective foil
keyboard and clean the remaining glued joints.
Glue the new foil keyboard precisely in the recess. Make sure the foil matches precisely
and the window of the foil keyboard lies exactly over the display.
Reassemble the control panel in reversed order and mount it back in the front housing.
Figure 37: Display fastening screws.