2-1
6600-0371-000 10/27/03
2/Setup and Checkout Procedures
2-1
In this section
2.1 Setup ................................................................................................................... 2-1
2.2 Mechanical Checkout Procedure ......................................................................... 2-1
A. Overall Appearance ......................................................................................... 2-1
B. Heater Assembly Rotation .............................................................................. 2-2
C. Mechanical Checks ........................................................................................ 2-2
D. Accessory Checks ......................................................................................... 2-2
E. Wall Mount checkout...................................................................................... 2-3
2.3 Control Unit Checkout Procedure ......................................................................... 2-4
A. Control Unit Check ......................................................................................... 2-4
B. Elapsed Timer Check ..................................................................................... 2-4
C. Observation Light Check ................................................................................. 2-5
D. Raise and Lower Bed Switch Check ............................................................... 2-5
D. Interlock Switch Check ................................................................................... 2-6
E. Power Failure, Memory and Battery Test........................................................ 2-6
2.1 Setup
Refer to the setup instructions shipped with the warmer for initial unpacking and setup of
the unit after shipment.
After removal from the shipping containers, inspect the Ohmeda Panda Baby Warmer and all
accessory items for any signs of damage which may have occurred during shipment. File a
damage claim with the shipping carrier if damage has occurred. Also confirm the presence of
all accessory items or factory installed options as listed on the packing slip.
2.2 Mechanical Checkout Procedure
WARNING: w Before using the Ohmeda Panda Baby Warmer, read this entire manual.
Attempting to use this device without a thorough understanding of its
operation may result in patient or user injury.
w Do not perform the Checkout Procedures (Mechanical and Control Unit)
while a patient occupies the warmer.
w Complete the “Checkout Procedures” section of this manual before putting
the unit into operation. If the warmer fails any portion of the checkout
procedures it must be removed from use and repaired.
A. Overall Appearance
1. Disconnect the power cord from the ac power source for the mechanical checks
portion of this procedure.
2. Check the overall appearance of the Panda Baby Warmer/ Bassinet System. There
should be no obvious damage.
3. Check that all casters are in firm contact with the floor and that the warmer is stable
and moves freely.
For the free standing warmer, check that all six casters on the warmer move freely.
Note: On the free steanding warmer, it is possible that the two center casters may not
be in contact with the floor at all times.