3/Operation
3-2
Elapsed Time:
The elapsed time display indicates elapsed time in minutes and seconds up to a maxi-
mum of 60 minutes.
B. Indicator Lights
• Servo Mode Indicator
• Manual Mode Indicator
• Apgar Tones Indicator
Servo Mode Indicator:
The servo mode indicator lights when the warmer is in the servo mode of operation.
For proper operation of the warmer in the Servo Mode see section 3.2.
Manual Mode Indicator:
The manual mode indicator lights when the warmer is in the manual mode of operation.
For proper operation of the warmer in the Manual Mode see section 3.2.
WARNING: w Use the servo mode unless the manual mode is specifically prescribed.
While both modes require patient monitoring, the manual mode requires
constant attention. In the manual mode, you must take the responsibility
for making heater adjustments or environmental changes in response to
the following conditions:
• patient size and weight;
• bed to heater head height differences;
• changes in the environment (drafts, direct sunlight, phototherapy lamp
usage, etc.);
• changes in the patient condition
In the servo mode, the warmer automatically adjusts heater output to
maintain the desired skin temperature, reducing (but not eliminating) the
need to monitor the patient and make environmental adjustments to
ensure adequate patient temperature.
Apgar Tones Indicator:
The Apgar tones indicator lights when the Apgar tones are activated, and the Apgar timer
is running.
C. Switches
•Mode Switch
• Increase (M) and Decrease (?) Switches
• Alarm Silence
• Start/Hold
• Reset
•Apgar Tones On/Off
• Light On/OffMode Switch
• Raise Bed (M) (Elevating models only)
• Lower Bed (?) (Elevating models only)