DIGITAL
ULTRASOUND
WATTMETER
UPM-DT-1PA, UPM-DT-10PA,
and UPM-DT-1000PA
USER MANUAL

Questions and Answers

  • J
    Julie ClineAug 16, 2025
    Why does the measurement fluctuate on my Ohmic instruments UPM-DT-1000PA?
    • R
      Rodney WilliamsAug 16, 2025
      The fluctuation of measurement in Ohmic instruments Measuring Instruments is caused by an unstable environment. The meter is very sensitive and is affected by minor air drafts and/or vibrations. To resolve this, place the meter on a solid surface and shield it from any air drafts.
  • M
    Madison WrightAug 19, 2025
    How to fix timeout on Ohmic instruments Measuring Instruments?
    • K
      Katherine JordanAug 19, 2025
      If you are experiencing a timeout error with your Ohmic instruments Measuring Instruments, it is due to an operation timeout. To resolve this, re-zero or re-tare the meter once the environment is stable.
  • J
    joeldunnAug 22, 2025
    Why does the measurement drift over time on Ohmic instruments UPM-DT-1000PA Measuring Instruments?
    • M
      Matthew ScottAug 22, 2025
      The measurement drift in Ohmic instruments Measuring Instruments could be due to several reasons: * The tank water has absorbed too much air. Replace the tank water with fresh, room temperature degassed water. * The conical target has developed bubbles on its surface. Wipe any bubbles from the surface of the conical target, including the bottom. * The transducer head has developed bubbles on its surface. Wipe any bubbles from the surface of the transducer head.
  • M
    Michael MeltonAug 25, 2025
    What to do if my Ohmic instruments UPM-DT-1000PA Measuring Instruments shows Err 8.3?
    • M
      Maria OwenAug 26, 2025
      If your Ohmic instruments Measuring Instruments displays 'Err 8.3', it indicates an over range situation. To resolve this, remove any excess force on the meter.
  • M
    Matthew RiceAug 27, 2025
    What to do if my Ohmic instruments UPM-DT-1000PA Measuring Instruments will not zero?
    • A
      Amber JohnsonAug 27, 2025
      If your Ohmic instruments Measuring Instruments is not zeroing, it could be due to the following reasons: * The degassed water has absorbed too much air. Refill with fresh degassed water. * The device requires calibration or repair. Contact Ohmic Instruments for calibration or repair.
  • T
    Tina CervantesAug 31, 2025
    What does Err 8.4 mean on Ohmic instruments UPM-DT-1000PA Measuring Instruments and how to fix it?
    • Z
      zfernandezAug 31, 2025
      The error code 'Err 8.4' on your Ohmic instruments Measuring Instruments indicates that the load was removed during measurement. To fix this, replace the load or transducer and repeat the test.
  • G
    Geoffrey NelsonSep 3, 2025
    Why does the Ohmic instruments UPM-DT-1000PA Measuring Instruments not respond to the test weight or transducer power?
    • I
      inewtonSep 3, 2025
      If your Ohmic instruments Measuring Instruments is not responding to the test weight or transducer power, it could be because the Conical Target Assembly and/or the Target Support Tube is rubbing or binding against its surrounding. Verify the Conical Target Assembly and its Support Tube are correctly positioned.
  • N
    Nicole WilsonSep 6, 2025
    What does 'Over init. Range' mean on Ohmic instruments Measuring Instruments?
    • A
      April JohnsonSep 6, 2025
      The 'Over init. Range' error on Ohmic instruments Measuring Instruments indicates that there was a load during power on. Remove the conical target assembly and cycle power to resolve this.
  • C
    Christopher MoyerSep 8, 2025
    What does 'Under init. range' mean on Ohmic instruments UPM-DT-1000PA?
    • E
      eleeSep 8, 2025
      The 'Under init. range' error on your Ohmic instruments Measuring Instruments means there was no load during power on. To fix this, disengage the slide lock on the back of the unit (if equipped).
  • K
    Kristina JohnsonSep 12, 2025
    What does 'Busy' mean on Ohmic instruments UPM-DT-1000PA?
    • M
      Manuel FoxSep 12, 2025
      The 'Busy' message on Ohmic instruments Measuring Instruments indicates that the device is currently performing an operation (tare, zero, printing, or waiting for a stable weight). Wait until operation completes.

Summary

Safety and Precautions

General Use and Modification Warnings

Warnings regarding intended use, unauthorized modifications, and service by qualified personnel.

Environmental and Handling Cautions

Cautions concerning operating environment, inspection for wear, and patient contact.

Liquid and Cleaning Cautions

Precautions against liquid spills, submersion, and proper cleaning methods for the device.

INTRODUCTION

TEST SETUP

METER LAYOUT

TESTING

PRINTING AND COMMUNICATIONS

TROUBLESHOOTING

Common Symptoms and Solutions

Addresses issues like fluctuating measurements, drifting, and zeroing problems with their causes and fixes.

Displayed Error Messages

Explains specific error codes (Timeout, Over Range, etc.) and their corresponding solutions.

SPECIFICATIONS

Ohmic instruments UPM-DT-1000PA Specifications

General IconGeneral
Accuracy±0.5% of reading ±0.2% of full scale
DisplayLCD
Resolution0.1 W
Power Supply100-240 V AC, 50/60 Hz
Operating Temperature0°C to 50°C
Power Range0 - 1000 W
OutputRS-232
Frequency Range45 Hz to 65 Hz