DIGITAL
ULTRASOUND
WATTMETER
UPM-DT-1PA, UPM-DT-10PA,
and UPM-DT-1000PA
USER MANUAL

Questions and Answers

  • T
    tiffany08Aug 15, 2025
    Why does the measurement fluctuate on my Ohmic instruments UPM-DT-10PA?
    • J
      James WolfeAug 15, 2025
      The fluctuation of measurement in Ohmic instruments Measuring Instruments is caused by an unstable environment. The meter is very sensitive and is affected by minor air drafts and/or vibrations. To resolve this, place the meter on a solid surface and shield it from any air drafts.
  • R
    Rebecca MillerAug 18, 2025
    How to fix timeout on Ohmic instruments UPM-DT-10PA?
    • C
      cheryl06Aug 18, 2025
      If you are experiencing a timeout error with your Ohmic instruments Measuring Instruments, it is due to an operation timeout. To resolve this, re-zero or re-tare the meter once the environment is stable.
  • B
    Brian ParkerAug 21, 2025
    Why does the measurement drift over time on Ohmic instruments UPM-DT-10PA Measuring Instruments?
    • B
      Blake WhiteAug 21, 2025
      The measurement drift in Ohmic instruments Measuring Instruments could be due to several reasons: * The tank water has absorbed too much air. Replace the tank water with fresh, room temperature degassed water. * The conical target has developed bubbles on its surface. Wipe any bubbles from the surface of the conical target, including the bottom. * The transducer head has developed bubbles on its surface. Wipe any bubbles from the surface of the transducer head.
  • T
    Tyler MartinAug 25, 2025
    What to do if my Ohmic instruments UPM-DT-10PA shows Err 8.3?
    • K
      Kevin CardenasAug 25, 2025
      If your Ohmic instruments Measuring Instruments displays 'Err 8.3', it indicates an over range situation. To resolve this, remove any excess force on the meter.
  • C
    Crystal AlexanderAug 27, 2025
    What to do if my Ohmic instruments Measuring Instruments will not zero?
    • A
      Amanda MurrayAug 27, 2025
      If your Ohmic instruments Measuring Instruments is not zeroing, it could be due to the following reasons: * The degassed water has absorbed too much air. Refill with fresh degassed water. * The device requires calibration or repair. Contact Ohmic Instruments for calibration or repair.
  • P
    Peter GuzmanAug 30, 2025
    What does Err 8.4 mean on Ohmic instruments UPM-DT-10PA Measuring Instruments and how to fix it?
    • M
      millerleslieAug 30, 2025
      The error code 'Err 8.4' on your Ohmic instruments Measuring Instruments indicates that the load was removed during measurement. To fix this, replace the load or transducer and repeat the test.
  • W
    William ThomasSep 2, 2025
    Why does the Ohmic instruments Measuring Instruments not respond to the test weight or transducer power?
    • J
      John JonesSep 2, 2025
      If your Ohmic instruments Measuring Instruments is not responding to the test weight or transducer power, it could be because the Conical Target Assembly and/or the Target Support Tube is rubbing or binding against its surrounding. Verify the Conical Target Assembly and its Support Tube are correctly positioned.
  • M
    Melanie PopeSep 5, 2025
    What does 'Over init. Range' mean on Ohmic instruments UPM-DT-10PA?
    • R
      robert99Sep 6, 2025
      The 'Over init. Range' error on Ohmic instruments Measuring Instruments indicates that there was a load during power on. Remove the conical target assembly and cycle power to resolve this.
  • J
    Joseph DudleySep 8, 2025
    What does 'Under init. range' mean on Ohmic instruments Measuring Instruments?
    • S
      Sarah PerkinsSep 8, 2025
      The 'Under init. range' error on your Ohmic instruments Measuring Instruments means there was no load during power on. To fix this, disengage the slide lock on the back of the unit (if equipped).
  • S
    Stephanie ThompsonSep 12, 2025
    What does 'Busy' mean on Ohmic instruments Measuring Instruments?
    • M
      Mary ZavalaSep 12, 2025
      The 'Busy' message on Ohmic instruments Measuring Instruments indicates that the device is currently performing an operation (tare, zero, printing, or waiting for a stable weight). Wait until operation completes.

Summary

Test Setup Procedures

Degassed Water Preparation

Details on degassing water, storage, temperature requirements, and testing recommendations.

Conical Target Installation

Transducer Placement Guide

Meter Layout and Controls

Front Panel Operation

Describes the function of buttons 1-4 on the front panel for power, printing, and mode selection.

Display and Stability Indicators

Explains the 7 segment display, selected unit, and stability indicators (items 5-7).

Testing the Ultrasound Wattmeter

Troubleshooting Common Issues

Product Specifications

Ohmic instruments UPM-DT-10PA Specifications

General IconGeneral
BrandOhmic instruments
ModelUPM-DT-10PA
CategoryMeasuring Instruments
LanguageEnglish