■ Connect a voltmeter as indicated and
adjust, using potentiometer p5 (ref. 1).
The output signal must be equal to
880 mV.
■ Next, inject the calibration gas at a
flow rate of 30 l/h (refer to paragraph
Indications for calibrating the CTX
300 SC, on page 53).
■ Wait for the signal to stabilize and
adjust the signal with the sensitivity
potentiometer p6 (ref. 2).
The output signal should be:
■ Stop injecting the calibration gas.
■ Verify that the reading returns to zero
(880 mV). If it does not, repeat the
entire procedure.
■ Calibration is complete.