28
CTX 300
User manual
Connect a voltmeter as indicated and
adjust, using potentiometer p5 (ref. 1).
The output signal must be equal to
880 mV.
Next, inject the calibration gas at a
flow rate of 30 l/h (refer to paragraph
Indications for calibrating the CTX
300 SC, on page 53).
Wait for the signal to stabilize and
adjust the signal with the sensitivity
potentiometer p6 (ref. 2).
The output signal should be:
Figure 31
U = 880 mV + 3520 mV x Cal gas concentration
Sensor measurement range
Example
Sensor measure
(% of full range)
Output signal
(mV)
0
880
50
2 640
100
4 400
Stop injecting the calibration gas.
Verify that the reading returns to zero
(880 mV). If it does not, repeat the
entire procedure.
Calibration is complete.
Flip the switch (ref. 1) into the MES
(Measure) position.
Figure 32