5 Maintenance
29
CTX 300 CO
2
calibration
This is a Carbone dioxide type sensor.
Warning: the sensor should be turned on for 15 minutes before
adjustments are made. The following text describes the steps
necessary to adjust the transmitter (first calibration).
Adjustment: if the current loop of the output signal has an impedance of
500 ohms, the power supply should never fall below 23 V DC.
Inject nitrogen at a rate of 30 l/h.
On the X1 terminal board, place an
ammeter between terminals 1 (+) and
2 (-).
With the potentiometer Z, adjust the
current to 4 mA.
With the ammeter still connected,
inject the test gas at a rate of 60 l/h.
Adjust the sensitivity with the
potentiometer S. Refer to the
paragraph Calibration curves, on
page 30 for sensors without
linearization cards.
If this fails, flip the J2 jumper and
begin again.