11 – Specific instructions
67
Special conditions of use
In case of exposure above the measuring range, it is mandatory to
bump test the instrument with gas and/or to perform a calibration.
In the event of a change of position, it is necessary to re-calibrate
the detector.
OLCT 100 IS (intrinsic safety mode of protection)
The detector must be powered by an intrinsically safe source.
The detector input characteristics on the J3 power plot are:
Ui = 28V
Ii = 93.3mA
Ci = 39.5nF
Li = 0H