Tool kit for maintenance.
Inject the calibration gas onto
the measurement sensor.
Impact on reading:
measurement similar to
measurement in diffusion mode.
Impact on response time: none.
6331141
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth
Used to take bypass readings.
Impact on reading: none if
calibration is performed under
the same conditions (pipe, flow
rate).
Impact on response time: none.
6327910
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth
Protects the detector from
liquids.
Impact on reading: none.
Impact on response time:
response time in diffusion mode
may increase for certain gases;
contact us for more information.
6329004
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth
Splash guard in
Stainless Steel
Protects the detector from
liquids.
Impact on reading: none.
Impact on response time:
response time in diffusion mode
may increase for certain gases;
contact us for more information.
Protects the detector from
liquids.
Impact on reading: none.
Impact on response time:
response time may increase for
certain gases; contact us for
more information.
6329014
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth
Remote gas
injection head
Used to detect ambient gases
while a calibration gas injection
pipe is being used. Only for
combustible gases,
Impact on reading: none.
Impact on response time:
negligible.
6327911
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth
Removable
PTFE protection
filter
Protects the gas inlet from
liquids and dust.
Impact on reading: none, but
this part cannot be used for the
detection of O
3
, HCL, HF and
Cl
2
.
Impact on response time:
response time may increase for
certain gases; contact us for
more information.
6335975
Plastic
material. Risk of
electrostatic
charges. Wipe
with a damp cloth