EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Electromagnetic immunity compliance information and
recommended electromagnetic environments
U
T
is the AC mains power supply prior to application of the
test level.
Immunity test
IEC 60601-1-2
test level
Compliance level Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:
±2, ±4, ±6 kV
Air:
±2, ±4, ±8 kV
Same as left Floors should by be made of wood, concrete,
or ceramic tile that hardly produces static. If
floors are covered with synthetic material that
tends to produce static, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV
for power supply lines
±1 kV
for input/output lines
Same as left Mains power quality should be that of a typical
commercial (original condition feeding the
facilities) or hospital environment.
Surge
IEC 61000-4-5
Differential mode:
±0.5, ±1 kV
Common mode:
±0.5, ±1, ±2 kV
Same as left Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
Same as left Mains power quality should be that of a typical
commercial or hospital environment. If the
user of this instrument requires continued
operation during power mains interruptions, it
is recommended that this instrument be
powered from an uninterruptible power supply
or a battery.
40% U
T
(60% dip in U
T
)
for 5 cycle
70% U
T
(30% dip in U
T
)
for 25 cycle
< 5% U
T
(> 95% dip in U
T
)
for 5 seconds
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Same as left It is recommended to use this instrument by
maintaining enough distance from any
equipment that operates with high current.