Appendix
103
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment such as cellular phones
Where “P” is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and “d” is the recommended separation
distance in meters (m).
This instrument complies with the requirements of
IEC 60601-1-2: 2001. However, under electromagnetic
environment that exceeds its noise level, electromagnetic
interference may occur on this instrument.
Electromagnetic interference may occur on this instrument
near a high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Immunity test
IEC 60601-1-2
test level
Compliance
level
Guidance
Formula for recommended separation distance
(V
1
=E
1
=3 according to the compliance level)
Conducted RF
IEC 61000-4-6
3Vrms
(150 kHz – 80 MHz)
3V (V
1
)
Radiated RF
IEC 61000-4-3
3V/m
(80 MHz – 2.5 GHz)
3V/m (E
1
)
80 MHz – 800 MHz
800 MHz – 2.5 GHz
d
3.5
V
1
---------
P
=
d
3.5
E
1
---------
P
=
d
7
E
1
-----
P
=