Chapter 4 Operation
63
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
To prevent the patient from accidentally biting the insertion
tube during an examination, it is strongly recommended that
a mouthpiece be placed in the patient’s mouth before
inserting the endoscope (for GIF models only).
To prevent the patient from accidentally loosing dental
prosthesis, make sure that the patient removes them before
the examination.
Do not apply olive oil or products containing petroleum-based
lubricants (e.g. vaseline). These products may cause
stretching and deterioration of the bending section’s
covering.
Do not allow the insertion tube to be bent within a distance of
10 cm or less from the junction of the boot. Insertion tube
damage can occur (see Figures 4.2 and 4.3).
Figure 4.2
Figure 4.3
> 10 cm
Endoscopes with flexibility adjustment
> 10 cm