Chapter 10 Troubleshooting
FLUSHING PUMP OFP-2
56
Problem Possible cause Action
Remote control function not
operating.
Remote control cable (MAJ-920)
fitted incorrectly.
Check connection according to MAJ-920
instructions.
Endoscope / video system center
function set incorrectly.
Check endoscope / video system center
settings according to manufacturer's
instructions.
Fluid continues to flow without
user activation.
Internal fault has occurred. Turn OFF the unit's power supply and
remove the tube from the endoscope or
EndoTherapy device.
Interference to other
equipment in the vicinity of the
unit.
Magnetic - proximity of sensitive
equipment to the pump motor’s
magnetic field.
Relocate sensitive equipment to a higher
shelf.
Electrical - typically distortion/
interference on monitor display
when equipment is operated.
Check system for damaged cables or
connectors.
Electromagnetic - interference
from other systems.
As above, plus investigate source of
interference.