USG-400, SONICBEAT & THUNDERBEAT
Troubleshooting Guide
Replace instrument: USG-400 IFU page 128-129
1 Without turning off the ultrasonic generator or disconnecting the transducer plug from the transducer socket of the
ultrasonic generator, disconnect the handle from the transducer and replace it with a new THUNDERBEAT or
SONICBEAT handle.
2 Probe check push button
Press the “PROBE CHECK” push button on the right side of the front panel.
The confirmation window (Probe Check) is displayed.
3 Executing the probe check
With the grasping section open, press and hold the handswitch or footswitch pedal for
three seconds and then release it.
• The probe check starts.
• The information window (Checking) is displayed.
• When the check is completed, the information window (Checking) disappears and the probe check
result is displayed.
Error window
(U504: Probe damage error)
Confirmation window
(Probe Check)