Table 12: Frequency Measurement/Source
Range Accuracy
(% of reading ± floor)
Read 2.0CPM - 600.0CPM 0.05% ± 0.1CPM
1.0Hz - 1000.0Hz 0.05% ± 0.1Hz
1.00KHz - 10.00KHz 0.05% ± 0.01KHz
Source 2.0CPM - 600.0CPM 0.05%
1.0Hz - 1000.0Hz 0.05%
1.00KHz - 10.00KHz 0.125%
Input voltage amplitude range on frequency is 1V to 20V zero based square wave
only.Output amplitude is adjustable from 1V to 20V, and is a square wave with 50% duty
cycle.For output frequency, a slight negative offset of approximately -0.1V is present to
assure zero crossing.
Table 13: Resistance Measurement
Range Accuracy
(% of reading ± floor)
Ohms low 0.00 - 400.0 0.015% ± 0.03
Ohms high 401.0 - 4000.0 0.015% ± 0.3
Table 14: Resistance Source
Range Excitation Current Accuracy
(% of reading ± floor)
Ohms low 5.0 - 400.0 0.1mA - 0.5mA 0.015% ± 0.1
5.0 - 400.0 0.5mA - 3mA 0.015% ± 0.03
Ohms high 400 - 1500 0.05mA - 0.8mA 0.015% ± 0.3
1500 - 4000 0.05mA - 0.4mA 0.015% ± 0.3
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is
<= 5ms.
Table 15: Thermocouple Measurement/Source
Range Accuracy
(% of reading ± floor)
Read (mV) -10.000mV - 75.000mV 0.02% ± 10µV
Source (mV) -10.000mV - 75.000mV 0.02% ± 10µV
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1